1000
Asia Pacific J Clin Nutr (1995) 4: 191-194
Asia Pacific J Clin Nutr (1995) 4: 191-194

Errors in determination of total
body protein by in vivo neutron activation of nitrogen due to non-uniform
neutron fluence inside the patient
Jukka Tölli, Lars Larsson and Magne Alpsten
Department of Radiation Physics, University
of Göteborg, Sahlgren Hospital, Göteborg, Sweden.
Total body protein can be estimated by in vivo neutron
activation of nitrogen. The method is based on capture of thermal
neutrons in a 14N(n,g )15N reaction. Sources
of error associated with this method, such as background subtraction,
variations in detection efficiency, etc, are analysed. Different
neutron reactions (absorption, elastic and inelastic scattering)
cause the neutron fluence to decrease inside the body. The activation
profile through the body is non-uniform which causes errors in the
calculation of total body nitrogen. A reduction of nitrogen by 5%
in a 3 cm thick volume near the body surface would result in an
error in the determination of total body nitrogen of approximately
0.3%. The error induced by changes in thickness of the subcutaneous
fat has also been estimated and the results show that a 5 mm change
in subcutaneous fat thickness changes the count rate from nitrogen
by 5%.
The experimental
arrangement
A 252CF source is used for in vivo neutron
activation analysis (IVNAA). The source is contained in a polyethylene
block which forms a collimator, surrounded by a 140 cm (diameter)
x 80 cm (height) water tank. The patients are irradiated from below
by a 15 cm x 50 cm neutron field. The distance between the neutron
source and patient is 70 cm.
The 10.8 MeV photons emitted in the reaction are detected
by two 15 cm x 15 cm NaI(TI)-detectors. A boron/plastic shield reduces
the neutron fluence inside the detectors. The reduction is 96%. The
detectors are also shielded with lead for reduction of the background
gamma radiation. The system is equipped with fast photomultiplier
tubes and an active type of dynode chains1. In this way
stability at high count rates has been obtained.
The pulses from each dynode chain is amplified by
two separate double delay line [DDL]-amplifiers. These amplifiers
are modified so that the 'delay lines' are removed and the they act
like fast amplifiers. The resulting pulse length at the output of
the dynode chain as well as from the fast amplifiers is less than
1m s. Each amplifier is connected to a fast ADC. The signals from both
ADCs are analysed in a fast multichannel analyser, which is connected
to a personal computer. The signal from each detector is handled and
analysed separately. The electronic system has been described more
in detail recently2. In this work hydrogen is not used
as internal standard as recommended by Vartsky et al. (1979)3
because of the high hydro 1000 gen background and the other reasons
quoted below.
Irradiation
geometry
In this work unilateral irradiation geometry has been
used with irradiation of the patient from below. Bilateral irradiation
is achieved by turning the patient through 180° half-way through the
measurement.
In the beginning of this project the detectors were
mounted side by side above the patient outside the primary neutron
field. This would partly compensate the lack of non-uniformity in
the neutron fluence inside the body. The detection efficiency of 10.8
MeV photons was determined for various depths using a 3 cm thick phantom
containing nitrogen at different depths in a water phantom. The results
from this experiment were compared to the results of a measurement
of thermal neutron fluence with In foils. The results of these two
measurements agree within 1 SD. Our conclusion is therefore that the
benefit of mounting the detectors above the patient is small.
In recent measurements the detectors have been positioned
beside the patient and in a plane perpendicular to the neutron flux.
It is much easier to shield the detectors in this position. There
are a number of other advantages in having the detectors placed in
this way. First, since the detectors can be shielded more effectively,
the count rate and the dead time decreases. The fact that the count
rate decreases gives a better signal-to-background ratio in the region
of interest. Second, the probability of pile-up events decreases which
gives a better energy resolution in all detected peaks in the pulse
height distribution. Thirdly, the mean energy of the neutrons which
reach the detector is lower, since they are scattered at least 90°
thus reducing the probability of neutron activation of the detectors
and surrounding materials.
Table 1 illustrates the improvements gained after
changing the detector positions. In this experiment phantoms of two
different sizes containing a 4.2% (by weight) solution of nitrogen,
were irradiated. As can he seen the count rate is decreased for both
phantoms.
Table 1. Improvements after change of detector
positions.
| Phantom Weight (kg)
|
Count rate (cps)
[9.7-11.3] MeV |
FWHM (keV)
at 2.23 MeV |
System dead time
(%) |
| 2.0 before |
13 800 |
167 |
11 |
| after |
3900 |
133 |
7 |
| 14.6 before |
9900 |
154 |
11 |
| after |
4500 |
134 |
9 |
We can also see that as the phantom size increases
the count rate decreases before the change by 28%, but after the change
the count rate increases by 15%. This result is due to the fact that
before a large phantom shielded the detectors more effectively than
small phantoms, which leads to a lower count rate. After the change
the count rate increases as a result of the increased amount of neutron
reactions in the irradiated object. The FWHM was measured at the hydrogen
peak at 2.23 MeV. The most important result is that the signal-to-background
ratio increased from 0.77 (old geometry) to 1.04 (new geometry).
Errors due
to non-uniform neutron fluence
Fast neutrons incident on the irradiated body are
slowed down by inelastic and elastic scattering reactions with the
body tissue until thermal equilibrium is reached. Once thermalized,
radiative capture causes the neutron fluence to decrease. A result
of this is a non-uniform neutron fluence through the irradiated body
and activation varying with depth inside the patient.
The fluence of thermal neutrons was determined by
activation of thin In foils in a water phantom. In Figure 1, thermal
neutron fluence is plotted versus depth in a water phantom for both
unilateral and bilateral irradiation geometry.
From this figure it is obvious that if the patient
gains protein near the body surface the result from IVNAA will overestimate
the true change in body nitrogen. In a similar way, if protein is
lost deeper in the body the IVNAA will under estimate the loss of
nitrogen.
The fluence in Figure 1 was divided in 20 sections;
the width of each section was 1 cm. The mean fluence through the body
is calculated by summation of each section. The mean fluence is a
measure of the probability of thermal neutron interactions in the
phantom. The percentage difference between the mean fluence and the
true fluence in each section was defined as the ratio:

The result from this calculation is shown in Figure
2.
The error resulting from the non-uniform neutron fluence
has been estimated by neutron activation of a large phantom containing
water, for background estimation. Thereafter the phantom was filled
with a known amount of nitrogen (urea). The net signal from nitrogen
reactions was used for calculation of a calibration factor. Into this
phantom a second phantom, formed as a 3 cm thick slice, containing
water, was introduced. Since the total amount of nitrogen is known
and constant, it is possible to estimate the relative error caused
by the second phantom at various depths. During irradiation the net
number of counts from nitrogen was measured. The error is calculated
as percentage difference between the measured nitrogen content and
the true nitrogen content. The result from this experiment is shown
in Figure 3. The solid line corresponds to the error in unilateral
geometry and the dotted line corresponds to bilateral geometry. It
should be noticed that this error represents a total loss of nitrogen
in the actual slice. As an example, if a patient loses nitrogen in
a region between 2 and 5 cm from body surface, a 100% loss in that
region would result in an error of 5.2%, while a 5% decrease of nitrogen
in that volume results in an error in the determination of TBN of
approxim 1000 ately 0.3% if unilateral irradiation geometry is assumed.
Figure 1. Thermal neutron fluence through patient.

Figure 2. Percentage difference between mean
and true neutron fluence.

Figure 3. Error due to non-uniform neutron
fluence.

Errors due changes in subcutaneous fat A common effect
in several diseases is a change in total body fat (TBF). In a case
where the thickness of body fat increases near the body surface the
estimated nitrogen content will be underestimated. As an example,
a common effect in patients treated with human growth hormone is that
TBN increases and TBF decreases4. In these patients the
measured change in TBN may partly be due to changes in body fat.
We have measured the effect of changes in the thickness
of subcutaneous fat by irradiation of a phantom containing a constant
amount of nitrogen. Layers of fat from a pig were placed between the
neutron source and nitrogen phantom. After subtraction of the background
the net numbers of counts from nitrogen were registered for different
depths of fat. In Figure 4 the change of nitrogen signal is plotted
versus fat thickness.
The results show that if the fat thickness decreases
by 5 mm the change in the nitrogen content will be overestimated by
approximately 5% if no correction is made.
Figure 4. Relative changes in nitrogen counts
due to changes in fat thickness.

Discussion
The problem with non-uniform activation/detection
profiles through the body is reduced in several other IVNAA facilities
by using the number of hydrogen counts as an internal standard5,6.
Hydrogen counts are used to reduce the corrections which are required
for differences in activation and detection efficiency arising from
differences in the body habitus. In this work the hydrogen counts
are not used in the calculation of TBN. The main reason for this is
that the polyethylene collimator contributes to the hydrogen signal
so much that the main part of the counts in the hydrogen signal is
from the collimator and not from the irradiated object. Another reason
is that if the hydrogen signal is used as an internal standard the
hydrogen and nitrogen must be distributed in a similar manner throughout
the body. In several diseases the patients may gain or loose extracellular
water which changes the relations between nitrogen and hydrogen.
The corrections required for differences in activation
and detection efficiency are minimized if the phantoms used for background
and calibration factors have the same size and shape as the patients.
It has been found that if the count rate is different in patient measurements
and the determination of calibration factors, the errors due to gain
or loss of pulse caused by random summing of gamma rays are not negligible7.
It is therefore important that the size of the phantom, used for calibration
factors, have approximately the same size and shape as the patient,
so that the measured overall count rates are equal in the phantom
and patient studies.
It should be noticed here that the errors do not cause
any problem if the patient gains or loses nitrogen uniformly. The
problem arises when the changes in the nitrogen content are non-uniform.
The resulting changes in the count rate from nitrogen
due c5a to changes in body fat can be corrected if the thickness of
body fat is measured.
References
- Kems, 1977.
- Larsson, 1992.
- Vartsky D, Prestwich WV, Thomas BJ, Dabek JT, Chettle
DR, Fremlin JH, Stammers K. The use of body hydrogen as an internal
standard in the measurement of nitrogen by prompt neutron capture
gamma rays. J Radioanal Chem 1979;48:243.
- Bengtsson B Å, Edén S, Lönn L, Kvist H, Stockland
A, Lindstedt G, Bosaeus 1, Tölli J, Sjöström L, Isaksson OGP Treatment
of adults with growth hormone (GH) deficiency with recombinant human
GH. J Clin Endocrin Metabol 1993; 76:309.
- Vartsky D, Ellis KJ, Cohn SH. In vivo measurement
of body nitrogen by analysis of prompt gamma rays from neutron capture.
J Nucl Med 1979; 20: 1158.
- Sutcliffe JE A correction for total body hydrogen
used as an internal standard in the measurement of total body protcin
by prompt gamma in vivo neutron activation analysis. Phys Mcd Biol
1989; 36:1681.
- Yankuba S, Oldroyd B, Metcalfe SC, Godfrey JE,
Oxby CB. Examination of errors caused by random summing in in vivo
prompt gamma analysis. Phys Med Biol 1989;34: 1089.
- Cohn SH, Dombrowski CS. Measurement of total body
calcium, sodium, chlorine and phosphorus in man by in vivo neutron
activation analysis. J Nucl Med 1971; 12:499.
- Harvey TC, Dykes PW, Chen NS, Ettinger KV, Jain
S, James H, Chettle DR, Fremlin JH, Thomas BJ. Measurement of whole
body nitrogen by neutron activation analysis. Lancet 1973:395.
- Leach MO, Thomas BJ, Vartsky D. Total body nitrogen
measured by the 14N(n,2n)13N method: a study
of the interfering reactions and the variation of spatial sensitivity
with depth. Int J Appl Rad Isot 1977; 28:263.
- McNeill KG, Borovnicar DJ, Krishman SS, Wang H,
Waana C, Harrison JE. Investigation of the factors which lead to
background in the measurement of nitrogen by IVNAA. Phys Med Biol
1989; 34:53.

Copyright © 1996 [Asia Pacific Journal of Clinical
Nutrition]. All rights reserved.
Revised:
January 19, 1999
.
to the top
0